Recommended optional strategic tests
Test Parameter |
Objective |
Maximum time interval | Test procedure* and key aspects |
Filter leakage |
Verify filter integrity |
12 months |
Filter media and filter seal integrity |
Containment leakage |
Verify absence of cross-contamination |
12 months |
Airflow direction and pressure differential |
Recovery (time) |
Verify clean-up time |
12 months |
Time taken maximum 15 minutes |
Airflow visualization |
Verify required airflow patterns |
12 months |
Airflow direction, documented evidence |
*Test procedure as per ISO 14644